| In order to improve the sensitivity
of the detector, the AURIGA collaboration is working on some R&D
projects focused on the redesign of the transducer-amplifier
chain. Obviously, before implementing the new devices on the detector they
must be tested in an experimental setup that mimics the final one: the
detector itself cannot be used as a transducer-amplifier
bench test!
To this end the AURIGA collaboration developed the Transducer
Test Facility (TTF) where to test readouts in a configuration
suitable for the integration
on the main detector. In particular the TTF should allow one to:
- operate the readout at the same temperature as it
would operate in AURIGA (approximately
0.1K). -> The TTF is thus equipped with a dilution
refrigerator.
- reduce mechanical (e.g. seismic) noise at 0.1K so
that it does not overcome the readout noise (e.g. transducer thermal
noise). -> An overall mechanical attenuation of at least -160dB is
required for frequencies around 1kHz (the frequency range of interest
for AURIGA).
- not to introduce spurious mechanical resonances
(e.g. due to the mechanical suspension) in the frequency range of
interest (i.e. 200÷1600Hz). -> Further constraints on the
suspension design.
- make fast (order of one week) thermal cycles
because readout optimization requires many cryogenic runs.
- have an experimental chamber wide enough to house
the readout (transducer + amplifier). Typical linear dimension must be
of the order of 30cm.
With these requirements in mind
the TTF was designed via a Finite Elements analysis and then machined.
Presently it appears as in the following 3-dim drawing:
Click
here for an enlarged image (110KB jpeg file).
Terms in the above picture are
explained and discussed in the following:
- Sylomer ®: soft rubber spring
with high internal friction: it provides the first room-temperature
isolation stage;
- Honeycomb:
very light and high strength material plate with internal overdamped
resonances;
- Dewar:
liquid Helium vessel that houses the experimental chamber;
- Room temperature suspension: TTF
support. It provides also for a mechanical isolation stage assuring
-65dB at about 1kHz;
- IVC: acronym for Internal Vacuum Chamber.
It is the experimental chamber that houses the last suspension stage
and the devices (e.g. transducer and amplifier) to be tested;
- Mixing
chamber: coldest part of the dilution refrigerator. The
measured lowest temperature measured in the TTF is 15mK (without any
thermal load).
- Low temperature suspension: the last
mechanical suspension stage. The upper part is connected by a
stainless steel spring to the IVC upper flange; the last annular mass
holds the transducer by means of copper springs that provide also for
the thermal link with the mixing chamber. The isolation is measured to
be -50dB at 1kHz for each spring: the inferred overall isolation is
thus -230dB at 1kHz.
|